APPLICATION
The Model 6900 is designed for the measurement of gas mixtures
produced in the industrial gas industry. The analyzer is designed
to quickly (3 - 5 minutes) analyze a gas cylinder, without the use
of calibration standards, wet chemicals, or special knowledge of
gas analysis instrumentation. The analyzer is designed so that it
may be used directly in the cylinder-filling environment.
The analyzer may also be used to check gas-mixing systems in the field. If desired. The analyzer may be run continuously as an on-stream analyzer. A linear 4-20 mA output is available on all ranges.
The analyzer use's Thermco's patented thermal conductivity detector/microprocessor technology to directly readout numerous mixtures created for the welding, heat treating, and food packaging industries.
The analyzer contains 9 ranges, which are selected with the push buttons on the front of the analyzer. All 9 ranges are factories calibrated. A custom ranges for calibration by the user us also available.
SPECIFICATIONS
RANGES
- 0-100% Carbon Dioxide in Argon
- 0-100% Oxygen in Argon
- 0-100% Helium in Argon
- 0-100% Hydrogen in Argon
- 0-100% Nitrogen in Argon
- 0-100% Carbon Dioxide in Nitrogen
- 0-100% Helium in Nitrogen
- 0-100% Hydrogen in Nitrogen
- Custom Range
ACCURACY
Valid over 50º F to 90º F(10ºC to 32ºC) ambient
temperature conditions.
±0.3% of minor component over range of 0-10.1%, all ranges.
Example: for a 5% O2 in Ar mixture, the analysis
will be between 4.7 and 5.3% O2.
±0.5% of minor component over a range of 10.1 % to 30.0%, all ranges.
Example: for a 25% CO2 in N2 mixture, the analysis
will be between 24.5 % and 25.5 % CO2.
±2% of minor component over a range of 60.1% to 100.0%, all ranges,
except CO2 in Ar, which has an accuracy of ±8%. Example: for a 75% He in Ar mixture, the analysis will be between 73.0% and 77.0% He.
MINIMUM/MAXIMUM OPERATING TEMP.
40ºFto 100ºF (4ºC to 38ºC)
RESOLUTION
0.1%
RESPONSE TIME
95% in 45 sec. for typical gas mixtures
UTILITIES REQUIRED
Reference Gas: Commercial grade argon or nitrogen (depending upon the base gas range used) is required to flow continuously at 0.5 SCFH during analysis.
Sample Gas: Moisture and oil free sample gas is required to flow continuously at 1.0 SCFH during analysis.
Sample, reference gas pressure: 1-100 PSIG
Power: 115 VAC, 60Hz , 1 Amp (220 VAC, 50/60 Hz. Optional)
FREQUENCY OF CALIBRATION
Calibration is recommended once a year when used under spot-checking conditions; if run continuously will depend on conditions and range.
WARM UP TIME
8 minutes under room temperature conditions
PHYSICAL SPECIFICATIONS
Size: 9". high, 11". wide, 14". deep (23 cm X 28
cm X 35.5 cm),
Weight: 22 pounds (10 kg)
ALTITUDE
Sea level to 2000 ft. (Altitude above 2000 ft. will cause some inaccuracy;
consult the factory)
ANALYZER OUTPUTS
Current: 4-20 mA, proportional to gas analysis range selected (available on all ranges). The 4-20 mA resolution equals display resolution of 0.1% gas analysis.